The PT-104 (-200 to 800 °C)
is a four (4) channel temperature measuring
data logger. It offers the ultimate in resolution (0.001 °C) and accuracy
(0.01 °C) when used with PT100 sensors. It can
also be used to measure
resistance and voltage.
A wide range of PT-100 sensors are available for use with the PT-104.
Although accurate temperature sensors are widely available, it has been difficult to take advantage of them due to errors caused by the measuring device. The PT104 however, is inherently accurate due to its novel design. Rather than relying on voltage references (which tend to be temperature sensitive) it uses 'reference' resistors which are extremely stable (low temperature co-efficient and drift). The exact value of each resistor is stored in an EEPROM to provide the ultimate in accuracy (yearly re-calibration is recommended). To achieve the 0.001 °C resolution a state of the art ADC is used that can resolve to better than 1 part in 16 million.
The PT104 measures temperature using platinum resistance thermometers (PRTs). Both common industry standards (PT100 and PT1000) are supported. The unit is compatible with 2, 3 and 4 wire sensors (4 wire PT100 sensors are recommended for accurate measurements). More information on sensors can be found below.
When measuring resistance, the unit uses a four-wire circuit to give the greatest possible accuracy.
Two resistance ranges are available (0 to 375 Ohms and 0 to 10,000 Ohms). The unit is calibrated for 0 to 375 Ohms so this range should be used for accurate measurements.
For voltage measurements, each input connector can be treated as a differential input with ground, or two single ended inputs. Both inputs must be zero Volts or above, though it does not matter which input has the higher voltage.
Two voltage ranges are available (0 to 115 mV and 0 to 2500 mV). For accurate measurements use the 0 to 2500 mV range.
The PT-104 is supplied with PicoLog data logging software for advanced data logging. PicoLog will automatically detect which sensor is connected and will display readings in the correct units. For users who wish to write their own custom software, drivers and examples are supplied.
The PT-104 is normally connected directly to a PC, but it is also possible to communicate with the PT-104 using a modem (radio or telephone) so that you can collect data from a remote site.
| Temperature | Resistance | Voltage | |
|---|---|---|---|
| Sensor | PT100*, PT1000 | N/A | N/A |
| Range | -200 to 800 °C | 0 to 375 Ω* 0 to 10 kΩ |
0 to 115 mV 0 to 2.5 V* |
| Linearity | 10 ppm | 10 ppm | 10 ppm |
| Accuracy @25 °C | 0.01 °C* | 20 ppm* | 0.2%* |
| Temperature coefficient | 3 ppm/°C | 3 ppm/°C | 100 ppm/°C |
| RMS Noise (using filter) | 0.01 °C | 10 ppm | 10 ppm |
| Resolution | 0.001 °C | 1 µΩ | 0.156 µV |
| Conversion time per channel | 720 mS ** | 720 mS ** | 180 mS |
| Number of inputs | 4 | ||
| Connectors | 4-pin miniDIN | ||
| Input impedence | >>1 MΩ | ||
| Overvoltage protection | ±100 V | ||
| Output | RS232, D9 female | ||
| Environmental | 20 to 30 °C for stated accuracy, 0 to 70 °C overall, 20 to 90% RH | ||
| Software |
PicoLog data
acquisition software for Windows 95/98/ME/NT/2000/XP. Drivers and examples C, pascal, Delphi, Visual Basic, Agilent VEE and LabVIEW. A macro is also provided to collect data directly into an Excel spreadsheet. |
||
| PRICES | PP161 Prices & Ordering | ||
| Quoted accuracy is for
options marked * ** For four wire temperature and resistance measurement |
|||